Keysight N5990A MIPI M-PHY Receiver Test User Guide
271
Sys Burst Tests
8
In
Compliance Mode
, the data generator sets the calibrated differential
voltage amplitude to the minimum (60mV for RT, 120mV for NT) and
maximum (245mV for RT, 490mV for NT) conformance values. The
common mode voltage is fixed to the nominal value (150mV) according to
the specification. No other signal impairments (ISI or jitter) are added.
At each step, the BER is measured and compared with the Target BER to
consider the pass/fail result.
In
Expert Mode
, the differential voltage component of each Level Pair is
tested over the given Amplitude Range. In this case, the result will show
the minimum and maximum differential voltage values at which the DUT
works properly and the test will pass if those values are within the spec
limits.
This test is performed at each selected lane.
Connection Diagram
Refer to the connection diagrams shown in the Connection Diagram
section of the Test 2.1.1- HS-RX Differential Input Voltage Amplitude Test
test.
Parameters
The parameters used in
Expert Mode
for this test, are as follows:
• CDR Loop Bandwidth (if CDR is selected in the Special Parameter
dialog, see
)
• CDR Peaking (if CDR is selected in the Special Parameter dialog, see
• CDR Transition Density (if CDR is selected in the Special Parameter
dialog, see
)
• CDR Loop Order (if CDR is selected in the Special Parameter dialog,
see
• Target BER
• IBerReader Init Mode
• Test Sequence
• Re-Init sequence after Reset DUT
• Lane Under Test Termination Model
• Level Pairs
• Amplitude Range
These parameters are listed in alphabetical order in
Summary of Contents for N5990A
Page 1: ...User Guide Keysight N5990A MIPI M PHY Receiver Test...
Page 62: ......
Page 86: ...86 Keysight N5990A MIPI M PHY Receiver Test User Guide 4 Calibrations...
Page 88: ...88 Keysight N5990A MIPI M PHY Receiver Test User Guide 4 Calibrations...
Page 214: ......
Page 234: ......
Page 268: ......
Page 298: ......