Keysight N5290/1A Installation Guide 4-3
Performance Verification
System Operator’s Check
System Operator’s Check
The operator’s check is a software driven test that checks the basic operation
of the assemblies in all of the measurement port signal paths. By performing
the operator’s check, the following are tested:
— attenuation ranges of all installed attenuators
— calibration of the receivers
— frequency response of the receivers
— phase lock and leveling
— basic functional test of noise floor and trace noise
Accessories Used in the Operator’s Check
Performing the Operator’s Check
1.
Press
UTILITY
>
System
>
Service
>
Verification
>
Operator’s Check
.
2.
In the PNA Operator’s Check dialog box (refer to
), under
Configure, select either:
— Prompt for attachment of Short/Open
, to pause at each step in the
process to allow moving the short/open to the appropriate port
— Shorts/Opens are attached to ALL ports
, to run through the test
without stopping. Shorts and opens can be mixed on the test ports
—
Additionally, you can choose:
—
Pause on Failure
—
Use unique log, to create a log file for this Operator’s
Check
3.
Click
Begin
.
4.
If shorts and opens are not connected to all ports, you will be prompted to
connect them as needed.
To achieve the maximum system stability, allow the analyzer to warm up
for at least 15 minutes before performing the Operator’s Check.
Equipment Type
Part Number
Female short, 1.0 mm
(any short from the 85059B calibration kits)
Female open, 1.0 mm
(any open from the 85059B calibration kits)