3- 22 Keysight N5290/1A Installation Guide
System Installation
Initializing the N5290/1A Broadband System
3-
Initializing the N5290/1A Broadband System
After the N5290/1A Broadband System has been calibrated and verified at the
factory, the PNA is reset back to "Standard PNA" mode to remove any artifact
files from the test process. To change your PNA over to "N529xA Broadband"
mode:
1.
On the PNA: press
Setup
>
External Hardware
>
Millimeter Config
2.
On the PNA: press
Instrument
>
Setup
>
External Hardware
>
Millimeter
Head Off
to
Millimeter Head On
.
3.
Power down the PNA.
4.
On the N5292A test set controller: verify the test set is powered down.
5.
Power up the PNA.
6.
Power up the N5292A.
The N5290/1A system should now initialize and recognize the mmW modules.
The port 1 module LED should be blue (addressed by the PNA and active) and
the other test port(s) should be green (no bias and on standby). See also,
“Troubleshooting the N5293/5A Millimeter Modules” on page 3-26
.