Service Information
2-
Measure Gain Factors for Ports 3 and 4 (Procedure B)
Repeat the process above for Ports 3 and 4 with the following modifications:
— Connect the SRC2 jumper on the rear panel.
— Connect to SRC2 IN in place of SRC1 IN.
— The first set of Test Set I/O commands should be 32,2 and 0,16.
(turn off SRC2 ALC and set the path to Port 3)
— The Test Set I/O command to set the path to Port 4 is 0,32.
Table 2-5
Gain Factor Addresses and Values
Ports/Freq
Write
Address
Read Address Value
Port 1 - 7 GHz
144
16
Port 1 - 10 GHz
148
20
Port 1 - 15 GHz
152
24
Port 1 - 20 GHz
156
28
Port 2 - 7 GHz
160
32
Port 2 - 10 GHz
164
36
Port 2 - 15 GHz
168
40
Port 2 - 20 GHz
172
44
Port 3 - 7 GHz
176
48
Port 3 - 10 GHz
180
52
Port 3 - 15 GHz
184
56
Port 3 - 20 GHz
188
60
Port 4 - 7 GHz
192
64
Port 4 - 10 GHz
196
68
Port 4 - 15 GHz
200
72
Port 4 - 20 GHz
204
76
Keysight N5261/62A User’s & Service Guide
81