7-102 Keysight N5247B Service Guide
Repair and Replacement Procedures
Post-Repair Procedures
7-
b.
When satisfied, click on
Modify
.
c.
Click
Exit.
Figure 7-50
Mechanical Cycle Counter window (Port 1 NOISE TUNER SWITCH and Port 2
NOISE RECVR SWITCH are only applicable for Option 029)
EEPROM Backup
The analyzer uses arrays of correction constants to enable the analyzer to
produce accurate, leveled source signals and receive clean test signals. These
constants are stored in non-volatile EEPROM memory and in flash memory
files.
The adjustments listed here generate new correction constants. The analyzer
must have a backup of this new data in case any of the data becomes
corrupted.
To store the backup data, perform these steps:
— Navigate to the EEPROM Backup Utility, located at: