6
Installation Note N5225-90105
Tools Required for the Installation
CAUTION
Use a 5/16-in torque wrench set to 10 in-lbs on all cable connections.
About Installing the Upgrade
Description
Qty
Part Number
T-10 TORX driver - set to 9 in-lbs (1.02 N.m)
1
N/A
T-20 TORX driver - set to 21 in-lbs (2.38 N.m)
1
N/A
5/16-in (8 mm) nutsetter or open end torque wrench - set to 10 in-lbs (1.13 N.m)
1
N/A
Products affected . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . N5224A and N5225A Option 400
Installation to be performed by . . . . . . . . . . . . . . . . . . . . . . Keysight service center or personnel qualified by
Keysight
Estimated installation time . . . . . . . . . . . . . . . . . . . . . . . . . 2 hours
Estimated adjustment time . . . . . . . . . . . . . . . . . . . . . . . . . 0.5 hour
Estimated full instrument calibration time . . . . . . . . . . . . . 4.5 hours
Summary of Contents for N5224AU-401