2-6 Keysight N5224A/25A Service Guide
General Product Information
Required Service Test Equipment
2-
5/16-inch, open-end torque
wrench
1.1 and 2.4 N-m
(10 and 21 in-lb) settings (for
semi-rigid cables)
N/A
N/A
T,R
1-inch, open-end torque
wrench
8.1 N-m (72 in-lb) setting (for
Port 1 and Port 2 connector nuts)
N/A
N/A
R
9-mm, socket or open-end
wrench
2.38 N-m (21 in-lb) setting (for
all front panel and most rear
panel connector hex nuts)
N/A
N/A
R
20-mm, open-end torque
wrench
0.9 N-m (8 in-lb) setting (for Port
1 and Port 2 measurement
connections)
N/A
N/A
R
Static Safety Parts
Adjustable antistatic wrist
strap
N/A
9300-1367
None
P,A,T
Antistatic wrist strap
grounding cord (5 foot)
N/A
9300-0980
None
P,A,T
Static control table mat and
earth ground wire
N/A
9300-0797
None
P,A,T
Miscellaneous
USB flash ROM drive
N/A
Any
None
P,A,R
a. Unless specified otherwise, equipment listed is required for all analyzer models.
b. P = Performance tests, A = Adjustments, T = Troubleshooting, R = Repair, V = System verification
Equipment
a
Critical Specifications
Recommended
Model or Part
Number
Alternate
Model Number
Use
b