3-44 Keysight N5290/1A Service Guide
System Tests and Adjustments
If the N5290/1A Operator’s Check Fails
3-
If the N5290/1A Operator’s Check Fails
1.
Clean the test ports, shorts, and adapters. Torque to specification. Repeat
the check.
2.
Run the adjustment tests and repeat the Operator’s Check. Refer
.
3.
If the check still fails, suspect a faulty component. Refer to
to begin troubleshooting to determine the
faulty component.
4.
Also, refer to