7-24 Keysight N5290/1A Service Guide
Repair and Replacement Procedures
Removing and Replacing the A1 Power Supply Assembly
7-
Removing and Replacing the A1 Power Supply Assembly
Tools Required
— T-10 TORX driver (set to 9 in-lb)
— T-20 TORX driver (set to 21 in-lb)
— ESD grounding wrist strap
Removal Procedure
Refer to
for this procedure.
1.
Disconnect the power cord.
2.
Remove the outer and inner covers. Refer to
.
3.
Using a T-10 TORX, remove the six pan head screws (item
①
) from the
power supply.
4.
Using a T-20 TORX, remove the pan head screw as indicated (item
➁
)
from the test set rear panel.
5.
Lift the A6 power supply assembly from the top of the instrument.
Replacement Procedure
1.
Reverse the order of the removal procedure.
2.
Perform the post-repair adjustments, verifications, and performance tests
that pertain to this removal procedure. Refer to
.