60 Installation Note N5222-90128
Installation Procedure for the Upgrade
-
— W36 (reuse) (N5222-20032) REF 1 RCVR R1 IN to A33 reference mixer
switch
* Refer to the note shown in
for information on installing W36
(part number N5222-20032).
Figure 37
Information to Secure W35 and W36 (N5222-20032, N5222-20033)
— W13 (reuse) (N5222-20005) A25 port 1 receiver coupler to A33 reference
mixer switch
* If the screws that attach the reference mixer switch to the test set deck
were loosened, torque these screws now to 9 in-lbs.
— W10 (reuse) (N5222-20043) W9 to A28 port 2 receiver coupler
— W8 (N5222-20042) W7 to A27 port 4 receiver coupler
— W6 (N5222-20041) W5 to A26 port 3 receiver coupler
— W4 (reuse) (N5222-20040) W3 to A25 port 1 receiver coupler
Install New Cable(s) – A2 1 HMA26.5 to A23/A24 Mixer Bricks
If your instrument has a new HMA26.5 (N5240-60101) installed: