![Keysight Technologies N4373D User Manual Download Page 119](http://html.mh-extra.com/html/keysight-technologies/n4373d/n4373d_user-manual_1962029119.webp)
Maintenance
Keysight
N4373D Lightwave Component Analyzer, User’s Guide
117
Alternative procedure
A better, more gentle, but more expensive
cleaning procedure is to use an ultrasonic bath with isopropyl
alcohol.
1 Hold the tip of the connector in the bath for at least three
minutes.
2 Take a new, dry soft tissue and remove the alcohol, dissolved
sediment and dust, by rubbing gently over the surface using a
small circular movement.
3 Blow away any remaining lint with compressed air.
How to clean connector interfaces
Preferred procedure
Use the following procedure on most
occasions.
1 Clean the interface by pushing and pulling a new, dry pipe
cleaner into the opening. Rotate the pipe cleaner slowly as
you do this.
2 Blow away any remaining lint with compressed air.
Procedure for stubborn dirt
Use this procedure when there is
greasy dirt on the interface.
1 Moisten a new pipe cleaner with isopropyl alcohol.
2 Clean the interface by pushing and pulling the pipe cleaner
into the opening. Rotate the pipe cleaner slowly as you do
this.
3 Moisten a new cotton swab with isopropyl alcohol.
4 Using a new, dry pipe cleaner, remove the alcohol, any
dissolved sediment and dust.
5 Blow away any remaining lint with compressed air.
How to clean bare fiber adapters
Bare fiber adapters are difficult to clean. Protect from dust
unless they are in use.
Be careful when using pipe cleaners, as the core and the bristles of
the pipe cleaner are hard and can damage the interface.
C A U T I O N
Summary of Contents for N4373D
Page 1: ...Keysight N4373D Lightwave Component Analyzer User s Guide ...
Page 2: ......
Page 3: ......
Page 38: ......
Page 84: ...User Interface Reference 82 Keysight N4373D Lightwave Component Analyzer User s Guide ...
Page 126: ......
Page 130: ......
Page 136: ......
Page 149: ...Specifications Keysight N4373D Lightwave Component Analyzer User s Guide 147 ...
Page 158: ...Specifications 156 Keysight N4373D Lightwave Component Analyzer User s Guide ...
Page 168: ......
Page 170: ...Ordering Information 168 Keysight N4373D Lightwave Component Analyzer User s Guide ...
Page 172: ......
Page 175: ......