Operating Information
54
N432A User’s Guide
Setting Measurement Limits
You can configure the N432A to detect when a measurement has crossed
over a predefined upper and/or lower limit value.
Limits are boundaries set for a certain power range and it can be applied
to the power, ratio, or difference measurement.
The N432A can be configured to verify the current measurement in any
measurement line against predefined upper and/or lower limit values. The
range of values that can be set for the upper and lower limits and the
default values depend on the measurement units in the currently selected
measurement line, as listed in the following table:
Table 2-2
Range of values for window limits
Procedure
The following procedure describes how to set the limits.
1
Press
>
to select the required window/measurement.
2
Use the up/down arrow key to highlight the
Limits
setting field.
3
Press
to select the
Limits
setting field.
4
Use the right arrow key to highlight the
Min Power
value field.
5
Press
to display the
Minimum Limit
pop- up (if the current unit is
dBm or dB) or the
Meter Minimum
pop- up (if the current unit is W or
%).
Window unit
Maximum
Minimum
Default maximum
Default minimum
dB
+200 dB
–180 dB
60 dB
–120 dB
dBm
+230 dBm
–150 dBm
90 dBm
–90 dBm
%
999.9 X%
100.0 a%
100.0 M%
100.0 p%
W
100.000 XW
1.000 aW
1.000 MW
1.000 pW
Meas Select
Summary of Contents for N432A
Page 2: ......
Page 3: ......
Page 20: ...XX N432A User s Guide ...
Page 22: ...XXII N432A User s Guide ...
Page 55: ...Operating Information 2 N432A User s Guide 33 4 Press to exit the channel setup screen ...
Page 90: ...2 Operating Information 68 N432A User s Guide Figure 2 36 Recall confirmation pop up dialog ...
Page 118: ...2 Operating Information 96 N432A User s Guide ...
Page 140: ...3 Maintenance 118 N432A User s Guide ...
Page 152: ...4 Specifications and Characteristics 130 N432A User s Guide ...
Page 153: ......