![Keysight Technologies N432A User Manual Download Page 144](http://html.mh-extra.com/html/keysight-technologies/n432a/n432a_user-manual_1962025144.webp)
4
Specifications and Characteristics
122
N432A User’s Guide
N432A Specifications
Frequency range
100 kHz to 18 GHz, thermistor- sensor dependent
Power range
–30 dBm to +10 dBm (1
μ
W to 10 mW), thermistor- sensor dependent
Thermistor sensor compatibility
•
Keysight 478A thermistor sensor (with Option H63, H75, and H76)
•
Keysight 8478B thermistor sensor
Meter power accuracy
Absolute power accuracy:
±
(0.1% of r 0.5
μ
W)
Meter voltage accuracies (1-year reference specifications)
•
V
RF
and V
COMP
: 23°C
±
5°C:
±
(0.0035% + 50
μ
V) [r range]
•
V
0
and V
1
: 23°C
±
5°C:
±
(0.0040% + 25
μ
V) [r range]
Bridge resistance
Selectable resistances of 100, 200, 300, and 400
Ω
Single sensor dynamic range
40 dB maximum (Keysight 478A and 8478B thermistor sensors)
Display units
•
Power: Absolute – Watts (W) or dBm, Relative – Percent (%) or dB
•
V
RF
and V
COMP
: VDC
•
V
0
and V
1
: VDC and mVDC
•
Bridge resistance: Ohm
Display resolutions
•
Power: Selectable resolutions of 1, 0.1, 0.01, and 0.001 dBm in
logarithmic mode, or 1, 2, 3, and 4 significant digits in linear mode
Summary of Contents for N432A
Page 2: ......
Page 3: ......
Page 20: ...XX N432A User s Guide ...
Page 22: ...XXII N432A User s Guide ...
Page 55: ...Operating Information 2 N432A User s Guide 33 4 Press to exit the channel setup screen ...
Page 90: ...2 Operating Information 68 N432A User s Guide Figure 2 36 Recall confirmation pop up dialog ...
Page 118: ...2 Operating Information 96 N432A User s Guide ...
Page 140: ...3 Maintenance 118 N432A User s Guide ...
Page 152: ...4 Specifications and Characteristics 130 N432A User s Guide ...
Page 153: ......