Keysight N4000/1/2A Operating and Service Manual
Figure 4-4
A typical ENR table after data entry
Operator’s check for T
cold
communication
The following procedure checks that the SNS can transfer the T
cold
data to the
NFA.
1
Remove any cables from the NFA input and SNS drive.
2
Press the
Preset
key. Wait until the preset routine is complete before
proceeding.
3
Press the
ENR
key.
4
Press the
Tcold
menu key.
NOTE
The preset condition used is the NFA default of
Preset (Factory)
.