N2750/1/2A User’s Guide
6
Performance Verification
Figure 49
Measurement Setup for V
out
22
On the VNA, press
[Trigger] > Single
to trigger a single sweep.
23
Press
[Scale] > Reference Level
and adjust the reference level until the 50 MHz
point (at the display’s left side) is at center screen.
NOTE
The reference level should be approximately –15.3 dB, but can vary a few
tenths of a dB either way.
24
The display should look similar to
NOTE
If it looks noticeably different, the probe tip wires may not be making contact
under the pinchers.