Using the Probes
1
N2750/1/2A User’s Guide
7
Oscilloscope Compatibility
The probes are compatible with 9000 Series, 90000 Series, or 90000 Q/X-Series
Infiniium oscilloscopes with software version 3.50 or above. Infiniium 90000
Q/X-Series requires the N5442A adapter. The probes are compatible with
3000X-series InfiniiVision oscilloscopes (one probe maximum connected to
oscilloscope). The probes are also compatible with 4000X-series InfiniiVision
oscilloscopes (up to four probes connected to oscilloscope).
Bandwidth
As shown in the following table, the available probe bandwidth depends on the
probe model, attached probe tip, and type of signal probed. To learn about the
supplied probe tips, refer to
Table 1
Available Probe Bandwidth
Signal Type
Browser Probe
Tip
Socketed Probe
Tip
Solder-In Probe
Tip
N2752A 6 GHz at 2:1 / 10:1
Differential
6 GHz
(7 GHz typical)
3 GHz
2.5 GHz
Single-Ended
6 GHz
(7 GHz typical)
1.2 GHz
2 GHz
Common-Mode
N/A
1 GHz
1.4 GHz
N2751A 3.5 GHz at 2:1 / 10:1
Differential
3.5 GHz
3 GHz
2.5 GHz
Single-Ended
3.5 GHz
1.2 GHz
2 GHz
Common-Mode
N/A
1 GHz
1.4 GHz
N2750A 1.5 GHz at 2:1 / 10:1
Differential
1.5 GHz
1.5 GHz
1.5 GHz
Single-Ended
1.5 GHz
1.2 GHz
1.5 GHz
Common-Mode
N/A
1 GHz
1.4 GHz
Summary of Contents for N2750A
Page 1: ...Keysight N2750 1 2A Differential Probes User s Guide ...
Page 28: ...28 N2750 1 2A User s Guide 1 Using the Probes ...
Page 34: ...34 N2750 1 2A User s Guide 2 SPICE Models Figure 16 SPICE Model ...
Page 38: ...38 N2750 1 2A User s Guide 2 SPICE Models ...
Page 42: ...42 N2750 1 2A User s Guide 3 Probe Calibration ...
Page 54: ...54 N2750 1 2A User s Guide 5 Performance Data Plots Figure 35 Frequency Response Common Mode ...
Page 56: ...56 N2750 1 2A User s Guide 5 Performance Data Plots Figure 38 Frequency Response Common Mode ...
Page 74: ...74 N2750 1 2A User s Guide 6 Performance Verification ...