Scale Fidelity
Procedure
Procedure
1.
Configure the equipment as shown in
.
2.
Preset the Source and press
Mode
,
Spectrum Analyzer
,
Mode Preset
on
the analyzer.
3.
Set up the synthesized sweeper by pressing:
Frequency
, 50,
MHz
Amplitude
, +5,
dBm
RF On/Off
, On
4.
Set up the signal analyzer by pressing:
Input/Output
,
More
,
Freq Ref In
,
External
FREQ Channel, Center Freq
, 50,
MHz
SPAN X Scale, Span
,
1, MHz
AMPTD Y Scale,
Ref Level
, 0,
dBm
Meas Setup, Average/Hold Number,
10,
Enter
Trace/Detector, Trace Average
Peak Search
5.
Set the external 10 dB step attenuator to 0 dB.
6.
Adjust the amplitude on the signal source until the marker amplitude on
the analyzer reads –15 dBm
±
0.2 dB.
7.
On the analyzer, press the
Single
,
Restart
to trigger a 10 sweep average.
8.
On the analyzer, activate the Marker Delta function by pressing
Peak
Search
,
Marker Delta
.
9.
Perform the following steps for each attenuator setting listed in the table
below:
a.
Select the next External attenuator setting.
b.
Press the
Restart
key to trigger a 10 sweep average.
c.
Enter the delta marker value into
d.
Check delta marker reading against the test limits.