14
Keysight M9195A/B PXIe Digital Stimulus/Response Startup Guide
Step 4: Install the Module
Module installation procedure
The module can be installed in any PXIe or hybrid PXI slot marked with a
peripheral slot compatibility image. Suitable slots will have a solid black circle
with a number inscribed. A PXIe slot is represented by a solid black circle (for
example,
); a hybrid PXI slot is represented by a solid black circle with the
letter H in the upper right (for example,
). For best performance, install the
module in a x8 hybrid slot; in the M9018A PXIe chassis, this is slot 2, 6, 11, or 15.
1
Make sure that the chassis line cord is plugged into an established earth
ground and that the chassis power switch is in the Off (Standby) position.
2
If the chassis has multiple fan speed settings, ensure that the fan switch is set
to AUTO.
3
Position the chassis to provide ample space between the chassis fan intake
and exhaust vents. Blockage by walls or obstructions affects the air flow
needed for cooling. (Refer to the chassis documentation for more information
about cooling).
4
If you are using an remote PC as the system controller, install the System
Interface Card in slot 1 of the chassis.
5
Before inserting the module into the chassis, back the mounting screws out to
ensure that there is no interference between the screws and the mounting
rails.
6
Holding the module by the injector/ejector handle, slide it into an available
PXI (or hybrid) slot, as shown in the figure below.
a
Install the module into the slot of the chassis by placing the module card
edges into the front module guides (top and bottom).
b
Slide the module to the rear of the chassis and ensure that the
injector/ejector handle is pushed down in the unlatched (downward)
position.