390
Keysight M8070A Programming Guide
5
SCPI Command Reference
:PLUGin:JTOLerance:STARt
Syntax
:PLUGin:JTOLerance:STARt ‘identifier’
Input
Parameters
‘identifier’: Specify the measurement name to start.
Description
This command starts a jitter tolerance measurement addressed by the
measurement name identifier.
This SCPI is applicable for M8041A, M8051A, M8061A, M8062A and
M8046A.
Example
The following example starts a jitter tolerance measurement addressed
with the measurement name identifier called ‘MyMeasurement’:
:PLUG:JTOL:STAR ‘MyMeasurement’
:PLUGin:JTOLerance:STOP
Syntax
:PLUGin:JTOLerance:STOP ‘identifier’
Input
Parameters
‘identifier’: Specify the measurement name to stop.
Description
This command stops a jitter tolerance measurement addressed by the
measurement name identifier.
This SCPI is applicable for M8041A, M8051A, M8061A, M8062A and
M8046A.
Example
The following example stops a jitter tolerance measurement with the
measurement name identifier called ‘MyMeasurement’:
:PLUG:JTOL:STOP ‘MyMeasurement’
:PLUGin:JTOLerance:STEP
Syntax
:PLUGin:JTOLerance:STEP ‘identifier’
Input
Parameters
‘identifier’: Specify the measurement name to step.
Description
This command allows you to step through a jitter tolerance measurement,
addressed by the measurement name identifier, that was stopped using
the BREak command.