Keysight M8070A Programming Guide
291
SCPI Command Reference
5
:PLUGin:ERATio:ACQuisition:TCLevel[?]
Syntax
:PLUGin:ERATio:ACQuisition:TCLevel ‘identifier’, <NRf>
:PLUGin:ERATio:ACQuisition:TCLevel? ‘identifier’
Input
Parameters
‘identifier’: Specify the measurement name.
<NRf>: Set target confidence level.
Return Range
0.1% to 99.9%
Description
This command sets the target confidence level when accumulation end is
set to pass/fail (in percent) for the addressed measurement name
identifier. The query returns the current target confidence level.
This SCPI is applicable for M8041A, M8051A, M8061A, M8062A and
M8046A.
Example
:PLUG:ERAT:ACQ:TCL ‘MyMeasurement’, 95
:PLUGin:ERATio:ACQuisition:TERatio[?]
Syntax
:PLUGin:ERATio:ACQuisition:TERatio ‘identifier’, <NRf>
:PLUGin:ERATio:ACQuisition:TERatio? ‘identifier’
Input
Parameters
‘identifier’: Specify the measurement name.
<NRf>: Set target error ratio.
Return Range
1E-18 to 1E-3
Description
This command sets the target error ratio for the addressed measurement
name identifier. The query returns the current target error ratio value.
This SCPI is applicable for M8041A, M8051A, M8061A, M8062A and
M8046A.
Example
:PLUG:ERAT:ACQ:TER ‘MyMeasurement’, 1e-09