242
Keysight M8070A Programming Guide
5
SCPI Command Reference
:INPut:COUPling[?]
Syntax
:INPut:COUPling ‘identifier’, <AC|DC>
:INPut:COUPling? ‘identifier’
Input
Parameters
‘identifier’: 'M*.DataIn1' or 'M*.DataIn2'
<AC|DC>: Select input coupling.
Return Range
AC|DC
Description
This command selects DC or AC input coupling.
This SCPI is applicable for M8041A and M8051A.
Example
:INP:COUP ‘M1.DataIn1’, DC
:INPut:SENSitivity[?]
Syntax
:INPut:SENSitivity ‘identifier’, <HIGH|NORMal>
:INPut:SENSitivity? ‘identifier’
Input
Parameters
‘identifier’: 'M*.DataIn1' or 'M*.DataIn2'
<HIGH|NORMal>: Select input coupling.
Return Range
HIGH|NORM
Description
This command sets the error detector input sensitivity to high or normal.
This SCPI is applicable for M8041A, M8051A and M8062A.
Example
:INP:SENS ‘M1.DataIn1’, HIGH
:INPut[:VOLTage]:WINDow[?]
Syntax
:INPut[:VOLTage]:WINDow[?] ‘Identifier’,< NRf>
:INPut[:VOLTage]:WINDow? ‘identifier’
Input
Parameters
‘identifier’: 'M*.DataIn'
Description
This command defines the input voltage window. The voltage applied at
the Data In connector needs to be within the voltage range defined by the
common mode voltage and the voltage window.