Keysight M8070A Programming Guide
209
SCPI Command Reference
5
:OUTPut:FREQuency:DIVider[?]
Syntax
:OUTPut:FREQuency:DIVider ‘identifier’, <NRf>
:OUTPut:FREQuency:DIVider? ‘identifier’
Input
Parameters
‘identifier’: 'M*.ClnClkOut’
Return Range
<NRf>
Description
This command divides the frequency at clean clock out (Cln Clk Out) on an
M8062A module. This SCPI is applicable for M8062A.
Example
:OUTPut:FREQuency:DIVider ‘M2.ClnClkout’, 2
:OUTPut:CALibration:SKEW[?]
Syntax
:OUTPut:CALibration:SKEW ‘identifier’, <NRf>
:OUTPut:CALibration:SKEW? ‘identifier’
Input
Parameters
‘identifier’: 'M*.DataOut1 or M*.DataOut2'
Return Range
-2ns to 2ns
Description
This command is used as offset to Data Delay and Jitter Delay.
This SCPI is applicable for M8041A, M8051A and M8045A.
Example
:OUTPut:CALibration:SKEW ‘M1.DataOut1’, 0
:OUTPut:EIDLe[:STATe][?]
Syntax
:OUTPut:EIDLe[:STATe] ‘identifier’, <OFF|IDLE|EXTernal>
:OUTPut:EIDLe[:STATe]? ‘identifier’
Input
Parameters
‘identifier’: 'M*.DataOut'
<OFF>:
Electrical idle feature is disabled.
<IDLE>:
Electrical idle feature is enabled (software controlled, power
saving mode).
<EXTernal>: Electrical idle input can be used to provide an electrical idle
signal at the Data Output. An external generated signal, for
example by a pulse generator at the electrical idle input,
controls the output stream at the Data Output.