174
Keysight M8070A Programming Guide
5
SCPI Command Reference
[:SOURce]:JITTer:LFRequency:RSSClocking:AMPLitude[?]
Syntax
[:SOURce]:JITTer:LFRequency:RSSClocking:AMPLitude ‘identifier’, <NRf>
[:SOURce]:JITTer:LFRequency:RSSClocking:AMPLitude? ‘identifier’
Input
Parameters
‘identifier’: 'M*.DataOut1', 'M*.DataOut2', M1.ClkOut’, or ‘M*.DataOut’
<NRf>: Set residual spread spectrum clocking amplitude.
Return Range
NRf
Description
This command sets the amplitude of the residual spread spectrum
clocking (rSSC). The units are set using the
[:SOURce]:JITTer:LFRequency:UNIT command.
This SCPI is applicable for M8041A, M8051A, M8061A and M8062A.
Example
:JITT:LFR:RSSC:AMPL ‘M2.DataOut2’, 0.016
[:SOURce]:JITTer:LFRequency:RSSClocking:FREQuency[?]
Syntax
[:SOURce]:JITTer:LFRequency:RSSClocking:FREQuency ‘identifier’, <NRf>
[:SOURce]:JITTer:LFRequency:RSSClocking:FREQuency? ‘identifier’
Input
Parameters
‘identifier’: 'M*.DataOut1', 'M*.DataOut2', ‘M1.ClkOut’, or ‘M*.DataOut’
<NRf>: Set residual spread spectrum clocking frequency.
Return Range
10 kHz to 100 kHz
Description
This command sets the frequency of the residual spread spectrum
clocking (rSSC). Acceptable units include Hz, kHz and exponents (for
example, 10E3 is the same as 10,000 Hz).
This SCPI is applicable for M8041A, M8051A, M8061A and M8062A.
Example
:JITT:LFR:RSSC:FREQ ‘M2.DataOut1’, 10khz