912
Keysight InfiniiVision M9241/42/43A PXIe Oscilloscopes SCPI Programmer's Guide
27
:SBUS<n> Commands
:SBUS<n>:UART Commands
NOTE
These commands are only valid when the UART/RS-232 triggering and serial decode option
has been licensed.
Table 112
:SBUS<n>:UART Commands Summary
Command
Query
Options and Query Returns
:SBUS<n>:UART:BASE
<base> (see
:SBUS<n>:UART:BASE?
<base> ::= {ASCii | BINary | HEX}
:SBUS<n>:UART:BAUDrat
e <baudrate> (see
:SBUS<n>:UART:BAUDrat
<baudrate> ::= integer from 100
to 8000000, 10000000, or 12000000
:SBUS<n>:UART:BITorde
r <bitorder> (see
:SBUS<n>:UART:BITorde
<bitorder> ::= {LSBFirst |
MSBFirst}
n/a
:SBUS<n>:UART:COUNt:E
<frame_count> ::= integer in NR1
format
:SBUS<n>:UART:COUNt:R
ESet (see
n/a
n/a
n/a
:SBUS<n>:UART:COUNt:R
XFRames? (see
<frame_count> ::= integer in NR1
format
n/a
:SBUS<n>:UART:COUNt:T
XFRames? (see
<frame_count> ::= integer in NR1
format
:SBUS<n>:UART:FRAMing
<value> (see
:SBUS<n>:UART:FRAMing
<value> ::= {OFF | <decimal> |
<nondecimal>}
<decimal> ::= 8-bit integer from
0-255 (0x00-0xff)
<nondecimal> ::= #Hnn where n ::=
{0,..,9 | A,..,F} for hexadecimal
<nondecimal> ::= #Bnn...n where n
::= {0 | 1} for binary
:SBUS<n>:UART:PARity
<parity> (see
:SBUS<n>:UART:PARity?
<parity> ::= {EVEN | ODD | NONE}
:SBUS<n>:UART:POLarit
y <polarity> (see
:SBUS<n>:UART:POLarit
<polarity> ::= {HIGH | LOW}