832
Keysight InfiniiVision M9241/42/43A PXIe Oscilloscopes SCPI Programmer's Guide
27
:SBUS<n> Commands
:SBUS<n>:LIN:TRIGger:SYMBolic:VALue
Command Syntax
:SBUS<n>:LIN:TRIGger:SYMBolic:VALue <data>
<data> ::= value in NR3 format
The :SBUS<n>:LIN:TRIGger:SYMBolic:VALue command specifies the signal value
to trigger on when LIN symbolic data has been loaded (recalled) into the
oscilloscope and the LIN trigger mode is set to FSIGnal.
Query Syntax
:SBUS<n>:LIN:TRIGger:SYMBolic:VALue?
The :SBUS<n>:LIN:TRIGger:SYMBolic:VALue? query returns the specified signal
value.
Return Format
<data><NL>
<data> ::= value in NR3 format
See Also
•
•
•
":SBUS<n>:LIN:TRIGger:SYMBolic:FRAMe"
•
":SBUS<n>:LIN:TRIGger:SYMBolic:SIGNal"
NOTE
Encoded signal values are not supported in the remote interface (even though they can be
used in the front panel graphical interface).