:MTESt Commands
23
Keysight InfiniiVision M9241/42/43A PXIe Oscilloscopes SCPI Programmer's Guide
563
:MTESt:AMASk:UNITs
Command Syntax
:MTESt:AMASk:UNITs <units>
<units> ::= {CURRent | DIVisions}
The :MTESt:AMASk:UNITs command alters the way the mask test subsystem
interprets the tolerance parameters for automasking as defined by
:MTESt:AMASk:XDELta and :MTESt:AMASk:YDELta commands.
• CURRent — the mask test subsystem uses the units as set by the
:CHANnel<n>:UNITs command, usually time for
Δ
X and voltage for
Δ
Y.
• DIVisions — the mask test subsystem uses the graticule as the measurement
system, so tolerance settings are specified as parts of a screen division. The
mask test subsystem maintains separate XDELta and YDELta settings for
CURRent and DIVisions. Thus, XDELta and YDELta are not converted to new
values when the UNITs setting is changed.
Query Syntax
:MTESt:AMASk:UNITs?
The :MTESt:AMASk:UNITs query returns the current measurement units setting for
the mask test automask feature.
Return Format
<units><NL>
<units> ::= {CURR | DIV}
See Also
•
"Introduction to :MTESt Commands"
•
•
•
•
•
Example Code
•