:MEASure Commands
21
Keysight InfiniiVision M9241/42/43A PXIe Oscilloscopes SCPI Programmer's Guide
509
:MEASure:STATistics:INCRement
Command Syntax
:MEASure:STATistics:INCRement
This command updates the statistics once (incrementing the count by one) using
the current measurement values. It corresponds to the front panel
Increment
Statistics
softkey in the Measurement Statistics Menu. This command lets you, for
example, gather statistics over multiple pulses captured in a single acquisition. To
do this, change the horizontal position and enter the command for each new pulse
that is measured.
This command is only allowed when the oscilloscope is stopped and quick
measurements are on.
The command is allowed in segmented acquisition mode even though the
corresponding front panel softkey is not available.
See Also
•
"Introduction to :MEASure Commands"
•
•
•
•