269
Keysight InfiniiVision M9241/42/43A PXIe Oscilloscopes
SCPI Programmer's Guide
9 :CALibrate Commands
Utility commands for viewing calibration status and for starting the user
calibration procedure. See
"Introduction to :CALibrate Commands"
Table 84
:CALibrate Commands Summary
Command
Query
Options and Query Returns
n/a
:CALibrate:DATE? (see
<return value> ::=
<year>,<month>,<day>; all in NR1
format
:CALibrate:LABel
<string> (see
:CALibrate:LABel?
<string> ::= quoted ASCII string
up to 32 characters
:CALibrate:OUTPut
<signal> (see
:CALibrate:OUTPut?
<signal> ::= {TRIGgers | MASK |
WAVEgen | WGEN1 | WGEN2 |
TSOurce}
Note: WAVE and WGEN1 are
equivalent.
Note: WGEN2 only available on
models with 2 WaveGen outputs.
n/a
:CALibrate:PROTected?
{"PROTected" | "UNPRotected"}
:CALibrate:STARt (see
n/a
n/a
n/a
:CALibrate:STATus?
<return value> ::=
<status_code>,<status_string>
<status_code> ::= an integer
status code
<status_string> ::= an ASCII
status string