:TRIGger Commands
31
Keysight InfiniiVision M9241/42/43A PXIe Oscilloscopes SCPI Programmer's Guide
1151
:TRIGger:TV:LINE
Command Syntax
:TRIGger:TV:LINE <line_number>
<line_number> ::= integer in NR1 format
The :TRIGger:TV:LINE command allows triggering on a specific line of video. The
line number limits vary with the standard and mode, as shown in the following
table.
Query Syntax
:TRIGger:TV:LINE?
The :TRIGger:TV:LINE? query returns the current TV trigger line number setting.
Return Format
<line_number><NL>
<line_number>::= integer in NR1 format
See Also
•
"Introduction to :TRIGger Commands"
•
•
Table 142
TV Trigger Line Number Limits
TV Standard
Mode
LINE
LFIeld1
LFIeld2
LALTernate
VERTical
NTSC
1 to 263
1 to 262
1 to 262
PAL
1 to 313
314 to 625
1 to 312
PAL-M
1 to 263
264 to 525
1 to 262
SECAM
1 to 313
314 to 625
1 to 312
GENERIC
1 to 1024
1 to 1024
1 to 1024
P480L60HZ
1 to 525
P720L60HZ
1 to 750
P1080L24HZ
1 to 1125
P1080L25HZ
1 to 1125
P1080L50HZ
1 to 1125
P1080L60HZ
1 to 1125
I1080L50HZ
1 to 1125
I1080L60HZ
1 to 1125