1136
Keysight InfiniiVision M9241/42/43A PXIe Oscilloscopes SCPI Programmer's Guide
31
:TRIGger Commands
:TRIGger:RUNT:QUALifier
Command Syntax
:TRIGger:RUNT:QUALifier <qualifier>
<qualifier> ::= {GREaterthan | LESSthan | NONE}
The :TRIGger:RUNT:QUALifier command selects the qualifier used for specifying
runt pulse widths:
• GREaterthan — triggers on runt pulses whose width is greater than the
:TRIGger:RUNT:TIME.
• LESSthan — triggers on runt pulses whose width is less than the
:TRIGger:RUNT:TIME.
• NONE — triggers on runt pulses of any width.
Query Syntax
:TRIGger:RUNT:QUALifier?
The :TRIGger:RUNT:QUALifier? query returns the runt trigger qualifier setting.
Return Format
<qualifier><NL>
<qualifier> ::= {GRE | LESS NONE}
See Also
•
"Introduction to :TRIGger Commands"
•
•