:TRIGger Commands
31
Keysight InfiniiVision M9241/42/43A PXIe Oscilloscopes SCPI Programmer's Guide
1081
:TRIGger:DELay:ARM:SOURce
Command Syntax
:TRIGger:DELay:ARM:SOURce <source>
<source> ::= CHANnel<n>
<n> ::= 1 to (# analog channels) in NR1 format
The :TRIGger:DELay:ARM:SOURce command selects the input used for the arming
edge in the Edge Then Edge trigger.
Query Syntax
:TRIGger:DELay:ARM:SOURce?
The :TRIGger:DELay:ARM:SOURce? query returns the current arming edge source.
Return Format
<source><NL>
<source> ::= CHAN<n>
See Also
•
"Introduction to :TRIGger Commands"
•
•
•