1034
Keysight InfiniiVision M9241/42/43A PXIe Oscilloscopes SCPI Programmer's Guide
29
:SYSTem Commands
:SYSTem:PRESet
Command Syntax
:SYSTem:PRESet
The :SYSTem:PRESet command places the instrument in a known state. This is the
same as pressing the
[Default Setup]
key or
[Save/Recall] > Default/Erase > Default
Setup
on the front panel.
When you perform a default setup, some user settings (like preferences) remain
unchanged. To reset all user settings to their factory defaults, use the *RST
command.
Reset conditions are:
Acquire Menu
Mode
Normal
Averaging
Off
# Averages
8
Analog Channel Menu
Channel 1
On
Channel 2
Off
Volts/division
5.00 V
Offset
0.00
Coupling
DC
Probe attenuation
10:1
Vernier
Off
Invert
Off
BW limit
Off
Impedance
1 M Ohm (cannot be changed)
Units
Volts
Skew
0
Cursor Menu
Source
Channel 1