Testing Performance
To verify bandwidth (-3 dB)
This test checks the bandwidth (-3 dB) of the oscilloscope. In this test you will use
a signal generator and a power meter.
1
Connect the equipment (see
a
Use the N cable to connect the signal generator to the input of the power
splitter input.
b
Connect the power sensor to one output of the power splitter.
c
Use an N-to-BNC adapter to connect the other splitter output to the
channel 1 input.
Table 11
Bandwidth (-3 dB) Test Limits
Models
Test Limits
6 GHz Models
All channels (-3 dB), dc to 6 GHz
4 GHz Models
All channels (-3 dB), dc to 4 GHz
2.5 GHz Models
All channels (-3 dB), dc to 2.5 GHz
1 GHz Models
All channels (-3 dB), dc to 1 GHz
Table 12
Equipment Required to Verify Bandwidth (-3 dB)
Equipment
Critical Specifications
Recommended Model/Part
Signal Generator
100 kHz - 6 GHz at 200 mVrms
Keysight N5171B with 6 GHz option
Power Meter
1 MHz - 6 GHz ±3% accuracy
Keysight N1914A
Power Sensor
1 MHz - 6 GHz ±3% accuracy
Keysight E9304A or N8482A
Power Splitter
outputs differ by < 0.20 dB
Keysight 11667A
Cable
Type N (m) 24 inch
Keysight 11500B
Adapter
Type N (m) to BNC (m)
Keysight 1250-0082 or
Pomona 3288 with Pomona 3533
Summary of Contents for InfiniiVision 6000 X Series
Page 1: ...Keysight InfiniiVision 6000 X Series Oscilloscopes Service Guide...
Page 12: ...12 6000 X Series Oscilloscopes Service Guide Tables...
Page 58: ...58 6000 X Series Oscilloscopes Service Guide 3 Calibrating and Adjusting...
Page 82: ...82 6000 X Series Oscilloscopes Service Guide 4 Troubleshooting...