16
6000 X-Series Oscilloscopes Service Guide
1
Characteristics and Specifications
Environmental Conditions
Specifications
Please see the
InfiniiVision 6000 X-Series Oscilloscopes Data Sheet
for complete,
up-to-date specifications and characteristics.
To download a copy of the data sheet please visit:
www.keysight.com/find/6000X-Series.
Contact us
To contact Keysight, see:
www.keysight.com/find/contactus
Environment
Indoor use only.
Ambient
temperature
Operating 0
°
C to +50
°
C; non-operating –30
°
C to +70
°
C
Humidity
Maximum Relative Humidity (non-condensing): 80% RH up to 40 °C, decreases
linearly to 50% at 50 °C. From 40 °C to 50 °C, the maximum % Relative Humidity
follows a line of constant dew point.
Altitude
Maximum operating altitude: 3,962 m (13,000 ft)
Overvoltage
Category
This product is intended to be powered by MAINS that comply to Overvoltage
Category II, which is typical of cord-and-plug connected equipment.
Pollution Degree
The InfiniiVision 6000 X-Series oscilloscopes may be operated in environments
of Pollution Degree 2 (or Pollution Degree 1).
Pollution Degree
Definitions
Pollution Degree 1: No pollution or only dry, non-conductive pollution occurs.
The pollution has no influence. Example: A clean room or climate controlled
office environment.
Pollution Degree 2. Normally only dry non-conductive pollution occurs.
Occasionally a temporary conductivity caused by condensation may occur.
Example: General indoor environment.
Pollution Degree 3: Conductive pollution occurs, or dry, non-conductive
pollution occurs which becomes conductive due to condensation which is
expected. Example: Sheltered outdoor environment.
Summary of Contents for InfiniiVision 6000 X Series
Page 1: ...Keysight InfiniiVision 6000 X Series Oscilloscopes Service Guide...
Page 12: ...12 6000 X Series Oscilloscopes Service Guide Tables...
Page 58: ...58 6000 X Series Oscilloscopes Service Guide 3 Calibrating and Adjusting...
Page 82: ...82 6000 X Series Oscilloscopes Service Guide 4 Troubleshooting...