:ACQuire Commands
7
Keysight InfiniiVision 3000T X-Series Oscilloscopes Programmer's Guide
239
:ACQuire:SEGMented:INDex
Command Syntax
:ACQuire:SEGMented:INDex <index>
<index> ::= an integer from 1 to 1000 (w/4M memory) in NR1 format
The :ACQuire:SEGMented:INDex command sets the index into the memory
segments that have been acquired.
The segmented memory acquisition mode is enabled with the :ACQuire:MODE
command. The number of segments to acquire is set using the
:ACQuire:SEGMented:COUNt command, and data is acquired using the :DIGitize,
:SINGle, or :RUN commands. The number of memory segments that have been
acquired is returned by the :WAVeform:SEGMented:COUNt? query. The time tag of
the currently indexed memory segment is returned by the
:WAVeform:SEGMented:TTAG? query.
The maximum number of segments may be limited by the memory depth of your
oscilloscope. For example, an oscilloscope with 1M memory allows a maximum of
250 segments.
Query Syntax
:ACQuire:SEGMented:INDex?
The :ACQuire:SEGMented:INDex? query returns the current segmented memory
index setting.
Return Format
<index><NL>
<index> ::= an integer from 1 to 1000 (w/4M memory) in NR1 format
See Also
•
•
•
•
•
•
•
•
•
"Introduction to :ACQuire Commands"
Example Code
' Segmented memory commands example.
' -------------------------------------------------------------------
NOTE
This command is available when the segmented memory option (Option SGM) is enabled.
Summary of Contents for InfiniiVision 3000T X Series
Page 1: ...Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide...
Page 40: ...40 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 1 What s New...
Page 50: ...50 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 2 Setting Up...
Page 64: ...64 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 3 Getting Started...
Page 254: ...254 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 8 BUS n Commands...
Page 306: ...306 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 12 DEMO Commands...
Page 364: ...364 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 17 FFT Commands...
Page 442: ...442 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 21 MARKer Commands...
Page 524: ...524 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 22 MEASure Commands...
Page 656: ...656 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 26 POWer Commands...
Page 666: ...666 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 27 RECall Commands...
Page 888: ...888 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 29 SBUS n Commands...
Page 978: ...978 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 30 SEARch Commands...
Page 1240: ...1240 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 38 Error Messages...