![Keysight Technologies InfiniiVision 3000T X Series Programmer'S Manual Download Page 1087](http://html1.mh-extra.com/html/keysight-technologies/infiniivision-3000t-x-series/infiniivision-3000t-x-series_programmers-manual_19623351087.webp)
:TRIGger Commands
33
Keysight InfiniiVision 3000T X-Series Oscilloscopes Programmer's Guide
1087
:TRIGger:ZONE<n>:VALidity
Query Syntax
:TRIGger:ZONE<n>:VALidity?
<n> ::= 1-2 in NR1 format
The :TRIGger:ZONE<n>:VALidity? query returns the validity of Zone 1 or Zone 2.
• INValid is returned (for Zone 2 only) when Zone 1 and Zone 2 overlap and have
opposing qualifying conditions (modes). Zone 1 can never be invalid.
• OSCReen (off-screen) is returned when the associated zone is off-screen, and
thus not being used to qualify the trigger.
• A zone is valid when it is neither invalid nor off-screen.
The validity of a zone is not affected by the zone's state. For example, a zone can
be valid and off. You cannot directly set the validity of a zone.
Return Format
<validity><NL>
<validity> ::= {VALid | INValid | OSCReen}
See Also
•
•
•
Summary of Contents for InfiniiVision 3000T X Series
Page 1: ...Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide...
Page 40: ...40 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 1 What s New...
Page 50: ...50 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 2 Setting Up...
Page 64: ...64 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 3 Getting Started...
Page 254: ...254 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 8 BUS n Commands...
Page 306: ...306 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 12 DEMO Commands...
Page 364: ...364 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 17 FFT Commands...
Page 442: ...442 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 21 MARKer Commands...
Page 524: ...524 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 22 MEASure Commands...
Page 656: ...656 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 26 POWer Commands...
Page 666: ...666 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 27 RECall Commands...
Page 888: ...888 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 29 SBUS n Commands...
Page 978: ...978 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 30 SEARch Commands...
Page 1240: ...1240 Keysight InfiniiVision 3000T X Series Oscilloscopes Programmer s Guide 38 Error Messages...