84
Keysight InfiniiVision 3000 X-Series Oscilloscopes User's Guide
4
Math Waveforms
•
Offset
— lets you set your own offset for the FFT. The offset value is in dB and
is represented by the center horizontal grid line of the display. See
adjust the math waveform scale and offset"
•
More FFT
— displays the More FFT Settings Menu.
2
Press the
More FFT
softkey to display additional FFT settings.
•
Window
— selects a window to apply to your FFT input signal:
•
Hanning
— window for making accurate frequency measurements or for
resolving two frequencies that are close together.
•
Flat Top
— window for making accurate amplitude measurements of
frequency peaks.
•
Rectangular
— good frequency resolution and amplitude accuracy, but use
only where there will be no leakage effects. Use on self-windowing
waveforms such as pseudo-random noise, impulses, sine bursts, and
decaying sinusoids.
•
Blackman Harris
— window reduces time resolution compared to a
rectangular window, but improves the capacity to detect smaller impulses
due to lower secondary lobes.
•
Vertical Units
— lets you select Decibels or V RMS as the units for the FFT
vertical scale.
•
Auto Setup
— sets the frequency Span and Center to values that will cause
the entire available spectrum to be displayed. The maximum available
frequency is half the FFT sample rate, which is a function of the time per
division setting. The FFT resolution is the quotient of the sampling rate and
the number of FFT points (f
S
/N). The current FFT Resolution is displayed
above the softkeys.
Summary of Contents for InfiniiVision 3000 X Series
Page 1: ...Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide...
Page 8: ...8 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide...
Page 24: ...24 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide...
Page 48: ...48 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 1 Getting Started...
Page 62: ...62 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 2 Horizontal Controls...
Page 72: ...72 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 3 Vertical Controls...
Page 122: ...122 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 6 Digital Channels...
Page 128: ...128 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 7 Serial Decode...
Page 134: ...134 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 8 Display Settings...
Page 182: ...182 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 11 Trigger Mode Coupling...
Page 200: ...200 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 12 Acquisition Control...
Page 210: ...210 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 13 Cursors...
Page 252: ...252 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 15 Mask Testing...
Page 256: ...256 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 16 Digital Voltmeter...
Page 290: ...290 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 19 Print Screens...
Page 324: ...324 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 21 Web Interface...
Page 344: ...344 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 22 Reference...