MIL-STD-1553/ARINC 429 Triggering and Serial Decode
27
Keysight InfiniiVision 3000 X-Series Oscilloscopes User's Guide
407
•
Line B (inverted)
.
•
Differential (A-B)
.
11
Press the
Auto Setup
softkey to automatically set these options for decoding and
triggering on ARINC 429 signals:
•
High Trigger Threshold: 3.0 V.
•
Low Trigger Threshold: -3.0 V.
•
Noise Reject: Off.
•
Probe Attenuation: 10.0.
•
Vertical Scale: 4 V/div.
•
Serial Decode: On.
•
Base: Hex.
•
Word Format: Label/SDI/Data/SSM.
•
Trigger: currently active serial bus.
•
Trigger Mode: Word Start.
12
If the high and low thresholds are not set correctly by
Auto Setup
:
•
Press the
High Threshold
softkey; then, turn the Entry knob to set the high
trigger threshold voltage level.
•
Press the
Low Threshold
softkey; then, turn the Entry knob to set the low
trigger threshold voltage level.
The threshold voltage levels are used in decoding and will become the trigger
levels when the trigger type is set to the selected serial decode slot.
ARINC 429 Triggering
To set up the oscilloscope to capture a ARINC 429 signal, see
After setting up the oscilloscope to capture a ARINC 429 signal:
1
Press
[Trigger]
.
2
In the Trigger Menu, press the
Trigger
softkey; then, turn the Entry knob to
select the serial slot (Serial 1 or Serial 2) on which the ARINC 429 signal is
being decoded.
Summary of Contents for InfiniiVision 3000 X Series
Page 1: ...Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide...
Page 8: ...8 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide...
Page 24: ...24 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide...
Page 48: ...48 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 1 Getting Started...
Page 62: ...62 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 2 Horizontal Controls...
Page 72: ...72 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 3 Vertical Controls...
Page 122: ...122 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 6 Digital Channels...
Page 128: ...128 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 7 Serial Decode...
Page 134: ...134 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 8 Display Settings...
Page 182: ...182 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 11 Trigger Mode Coupling...
Page 200: ...200 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 12 Acquisition Control...
Page 210: ...210 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 13 Cursors...
Page 252: ...252 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 15 Mask Testing...
Page 256: ...256 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 16 Digital Voltmeter...
Page 290: ...290 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 19 Print Screens...
Page 324: ...324 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 21 Web Interface...
Page 344: ...344 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 22 Reference...