I2C/SPI Triggering and Serial Decode
25
Keysight InfiniiVision 3000 X-Series Oscilloscopes User's Guide
379
2
In the Search Menu, press the
Search
softkey; then, turn the Entry knob to select
the serial slot (Serial 1 or Serial 2) on which the I2C signal is being decoded.
3
Press
Search
; then, select from these options:
•
Missing Acknowledge
— finds SDA data is high during any Ack SCL clock bit.
•
Address with no Ack
— finds when the acknowledge of the selected address
field is false. The R/W bit is ignored.
•
Restart
— finds when another start condition occurs before a stop condition.
•
EEPROM Data Read
— finds EEPROM control byte value 1010xxx on the SDA
line, followed by a Read bit and an Ack bit. It then looks for the data value
and qualifier set by the Data is softkey and the Data softkeys.
•
Frame(Start:Address7:Read:Ack:Data)
— finds a read frame on the 17th clock
edge if all bits in the pattern match.
•
Frame(Start:Address7:Write:Ack:Data)
— finds a write frame on the 17th clock
edge if all bits in the pattern match.
•
Frame(Start:Address7:Read:Ack:Data:Ack:Data2)
— finds a read frame on the 26th
clock edge if all bits in the pattern match.
•
Frame(Start:Address7:Write:Ack:Data:Ack:Data2)
— finds a write frame on the
26th clock edge if all bits in the pattern match.
For more information on searching data, see
For more information on using the
[Navigate]
key and controls, see
Setup for SPI Signals
Serial Peripheral Interface (SPI) signals setup consists of connecting the
oscilloscope to a clock, MOSI data, MISO data, and framing signal, then setting
the threshold voltage level for each input channel, and finally specifying any other
signal parameters.
To set up the oscilloscope to capture SPI signals, use the
Signals
softkey which
appears in the Serial Decode Menu:
1
Press
[Label]
to turn on labels.
2
Press
[Serial]
.
3
Press the
Serial
softkey, turn the Entry knob to select the desired slot (Serial 1 or
Serial 2), and press the softkey again to enable decode.
Summary of Contents for InfiniiVision 3000 X Series
Page 1: ...Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide...
Page 8: ...8 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide...
Page 24: ...24 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide...
Page 48: ...48 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 1 Getting Started...
Page 62: ...62 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 2 Horizontal Controls...
Page 72: ...72 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 3 Vertical Controls...
Page 122: ...122 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 6 Digital Channels...
Page 128: ...128 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 7 Serial Decode...
Page 134: ...134 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 8 Display Settings...
Page 182: ...182 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 11 Trigger Mode Coupling...
Page 200: ...200 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 12 Acquisition Control...
Page 210: ...210 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 13 Cursors...
Page 252: ...252 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 15 Mask Testing...
Page 256: ...256 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 16 Digital Voltmeter...
Page 290: ...290 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 19 Print Screens...
Page 324: ...324 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 21 Web Interface...
Page 344: ...344 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 22 Reference...