270
Keysight InfiniiVision 3000 X-Series Oscilloscopes User's Guide
17
Waveform Generator
1
Press the
Type
softkey and turn the Entry knob to select
Frequency Modulation
(FM)
.
2
Press the
Waveform
softkey and turn the Entry knob to select the shape of the
modulating signal:
•
Sine
•
Square
•
Ramp
•
Sine Cardinal
•
Exponential Rise
•
Exponential Fall
When the
Ramp
shape is selected, a
Symmetry
softkey appears so that you can
specify the amount of time per cycle that the ramp waveform is rising.
3
Press the
FM Freq
softkey and turn the Entry knob to specify the frequency of the
modulating signal.
4
Press the
FM Dev
softkey and turn the Entry knob to specify the frequency
deviation from the original carrier signal frequency.
When the modulating signal is at its maximum amplitude, the output frequency
is the carrier signal frequency plus the deviation amount, and when the
modulating signal is at its minimum amplitude, the output frequency is the
carrier signal frequency minus the deviation amount.
The frequency deviation cannot be greater than the original carrier signal
frequency.
Also, the sum of the original carrier signal frequency and the frequency
deviation must be less than or equal to the maximum frequency for the selected
waveform generator function plus 100 kHz.
The following screen shows an FM modulation of a 100 kHz sine wave carrier
signal.
Summary of Contents for InfiniiVision 3000 X Series
Page 1: ...Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide...
Page 8: ...8 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide...
Page 24: ...24 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide...
Page 48: ...48 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 1 Getting Started...
Page 62: ...62 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 2 Horizontal Controls...
Page 72: ...72 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 3 Vertical Controls...
Page 122: ...122 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 6 Digital Channels...
Page 128: ...128 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 7 Serial Decode...
Page 134: ...134 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 8 Display Settings...
Page 182: ...182 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 11 Trigger Mode Coupling...
Page 200: ...200 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 12 Acquisition Control...
Page 210: ...210 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 13 Cursors...
Page 252: ...252 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 15 Mask Testing...
Page 256: ...256 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 16 Digital Voltmeter...
Page 290: ...290 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 19 Print Screens...
Page 324: ...324 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 21 Web Interface...
Page 344: ...344 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 22 Reference...