262
Keysight InfiniiVision 3000 X-Series Oscilloscopes User's Guide
17
Waveform Generator
Creating New Arbitrary Waveforms
The New Waveform Menu is opened by pressing
Create New
in the Edit Waveform
Menu.
To create a new arbitrary waveform:
1
In the New Waveform Menu, press
Initial Pts
; then, use the Entry knob to select
the initial number of points in the new waveform.
The new waveform will be a square wave with the number of points you specify.
The points are evenly spaced over the time period.
2
Use the
Frequency/Frequency Fine/Period/Period Fine
softkey to set the time period
bounding parameter (repetition frequency) of the arbitrary waveform.
Edit Existing
Opens the Edit Waveform Points Menu. See
"Editing Existing Arbitrary Waveforms"
Interpolate
Specifies how lines are drawn between arbitrary waveform points.
When enabled, lines are drawn between points in the waveform editor. Voltage levels change
linearly between one point and the next.
When disabled, all line segments in the waveform editor are horizontal. The voltage level of
one point remains until the next point.
Source
Selects the analog channel or reference waveform to be captured and stored to the arbitrary
waveform. See
"Capturing Other Waveforms to the Arbitrary Waveform"
Store Source to Arb
Captures the selected waveform source and copy it to the arbitrary waveform. See
"Capturing Other Waveforms to the Arbitrary Waveform"
Softkey
Description
NOTE
You can use the
[Save/Recall]
key and menu to save arbitrary waveforms to one of four
internal storage locations or to a USB storage device, and you can recall them later. See
Summary of Contents for InfiniiVision 3000 X Series
Page 1: ...Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide...
Page 8: ...8 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide...
Page 24: ...24 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide...
Page 48: ...48 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 1 Getting Started...
Page 62: ...62 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 2 Horizontal Controls...
Page 72: ...72 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 3 Vertical Controls...
Page 122: ...122 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 6 Digital Channels...
Page 128: ...128 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 7 Serial Decode...
Page 134: ...134 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 8 Display Settings...
Page 182: ...182 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 11 Trigger Mode Coupling...
Page 200: ...200 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 12 Acquisition Control...
Page 210: ...210 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 13 Cursors...
Page 252: ...252 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 15 Mask Testing...
Page 256: ...256 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 16 Digital Voltmeter...
Page 290: ...290 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 19 Print Screens...
Page 324: ...324 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 21 Web Interface...
Page 344: ...344 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 22 Reference...