Triggers
10
Keysight InfiniiVision 3000 X-Series Oscilloscopes User's Guide
165
1
Press the
Time >
softkey; then, turn the Entry knob to set the time to
greater-than the sync-pulse width so that the oscilloscope synchronizes to the
vertical sync.
2
Press the
Edge #
softkey; then, turn the Entry knob to select the Nth edge after
after vertical sync to trigger on.
3
To enable or disable the horizontal sync control, press the first
Horiz Sync
softkey.
•
For interleaved video, enabling the
Horiz Sync
control and setting the
Horiz Sync
adjustment to the sync time of the probed video signal allows the
Edge #
function to count only lines and not double count during equalization.
Additionally, the
Field Holdoff
can be adjusted so that the oscilloscope
triggers once per frame.
•
Similarly, for progressive video with a tri-level sync, enabling the
Horiz Sync
control and setting the
Horiz Sync
adjustment to the sync time of the probed
video signal allows the
Edge #
function to count only lines and not double
count during vertical sync.
When the horizontal sync control is enabled, press the second
Horiz Sync
softkey; then, turn the Entry knob to set the minimum time the horizontal sync
pulse must be present to be considered valid.
To trigger on a specific line of video
Video triggering requires greater than 1/2 division of sync amplitude with any
analog channel as the trigger source. Turning the trigger
Level
knob in Video
trigger does not change the trigger level because the trigger level is automatically
set to the sync pulse tips.
One example of triggering on a specific line of video is looking at the vertical
interval test signals (VITS), which are typically in line 18. Another example is
closed captioning, which is typically in line 21.
1
Press the
[Trigger]
key.
2
In the Trigger Menu, press the
Trigger
softkey; then, turn the Entry knob to
select
Video
.
3
Press the
Settings
softkey, then press the
Standard
softkey to select the
appropriate TV standard (NTSC).
Summary of Contents for InfiniiVision 3000 X Series
Page 1: ...Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide...
Page 8: ...8 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide...
Page 24: ...24 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide...
Page 48: ...48 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 1 Getting Started...
Page 62: ...62 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 2 Horizontal Controls...
Page 72: ...72 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 3 Vertical Controls...
Page 122: ...122 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 6 Digital Channels...
Page 128: ...128 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 7 Serial Decode...
Page 134: ...134 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 8 Display Settings...
Page 182: ...182 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 11 Trigger Mode Coupling...
Page 200: ...200 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 12 Acquisition Control...
Page 210: ...210 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 13 Cursors...
Page 252: ...252 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 15 Mask Testing...
Page 256: ...256 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 16 Digital Voltmeter...
Page 290: ...290 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 19 Print Screens...
Page 324: ...324 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 21 Web Interface...
Page 344: ...344 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 22 Reference...