Calibrating the Oscilloscope
5
Keysight Infiniium MXR-Series Real-Time Oscilloscopes User's Guide
63
Calibration Procedure
1
Let the oscilloscope warm up before running the calibration.
2
Click
Utilities > Calibration...
.
3
Clear the
Cal Memory Protect
check box.
You cannot run a user calibration if this check box is selected. See the following
figure.
4
Click
Start Frame Cal
; then, follow the instructions on the screen.
The routine will prompt you to follow these steps:
a
Disconnect everything from all inputs and Aux Out.
b
Select the level of calibration you want to perform.
NOTE
Performing a user calibration will invalidate the oscilloscope's Certificate of Calibration. If
NIST (National Institute of Standards and Technology) traceability is required, perform the
, “Testing Performance,” starting on page 65 using traceable
sources.
CAUTION
The oscilloscope must be warmed up (with the oscilloscope application running) for at
least 30 minutes at ambient temperature before starting the calibration procedure.
Failure to allow warm up may result in inaccurate calibration.
Clear this
check box
Summary of Contents for Infiniium MXR Series
Page 1: ...Keysight Infiniium MXR Series Real Time Oscilloscopes User s Guide...
Page 10: ...10 Keysight Infiniium MXR Series Real Time Oscilloscopes User s Guide Index...
Page 20: ...20 Keysight Infiniium MXR Series Real Time Oscilloscopes User s Guide 1 Setting Up...
Page 106: ...106 Keysight Infiniium MXR Series Real Time Oscilloscopes User s Guide 6 Testing Performance...