Keysight Infiniium MXR-Series Real-Time Oscilloscopes User's Guide
2
Getting Started
Connecting Oscilloscope Probes
The Keysight Infiniium MXR-Series oscilloscopes are not rated for Measurement
Category II, III, or IV.
For complete documentation on Keysight oscilloscope probes for your Infiniium
oscilloscope, see the Probe Resource Center at:
To connect oscilloscope probes:
1
Attach the probe connector to the desired oscilloscope channel or trigger input
using the probe instructions.
2
Connect the probe to the circuit of interest using the browser or other probing
accessories.
Figure 9
MXR-Series oscilloscope probe connectors
~
~
+
+
8 or 4 analog input channels where probes are connected
CAUTION
Maximum input voltage at analog inputs
Do not exceed the maximum input voltage rating.
The maximum input voltage for the 50
Ω
input impedance setting is ±5 V.
The maximum input voltage for the 1 M
Ω
input impedance setting is 30 Vrms or
±40 Vmax (DC+Vpeak).
Probing technology allows for testing of higher voltages; the included N2873A 10:1
probe supports 300 Vrms or ±400 Vmax (DC+Vpeak). No transient overvoltage
allowed. Mains isolated circuits only.
CAUTION
When measuring voltages over 30 V, use a 10:1 probe.
Summary of Contents for Infiniium MXR Series
Page 1: ...Keysight Infiniium MXR Series Real Time Oscilloscopes User s Guide...
Page 10: ...10 Keysight Infiniium MXR Series Real Time Oscilloscopes User s Guide Index...
Page 20: ...20 Keysight Infiniium MXR Series Real Time Oscilloscopes User s Guide 1 Setting Up...
Page 106: ...106 Keysight Infiniium MXR Series Real Time Oscilloscopes User s Guide 6 Testing Performance...