Replacing Assemblies
Removing and replacing front panel assembly parts
To replace any of the front panel assembly parts, first remove the front panel
assembly as described in the previous section.
Removing front panel clutches, bezel, and keyboard
1
Use a thin item such as a ribbon to protect the oscilloscope from scratches, and
pull on the dark gray plastic clutches to pop them off the oscilloscope.
2
The front panel bezel has two tabs on each side, one tab on top, and one tab on
the bottom. Pop these tabs with your fingers to dislodge the front panel bezel.
When re-attaching the front panel bezel, carefully pull out the top and bottom
tabs slightly to prevent damaging the tabs.
3
Disconnect the cables and remove the two screws shown below from the board.
WARNING
This step will ruin the clutches. You will need to replace them with new ones.