88
Infiniium V-Series Oscilloscopes Service Guide
5
Replacing Assemblies
Removing and replacing the top panel and bottom panel
1
Disconnect the power cable, all probes and BNC input cables, and any other
cables including those for a mouse, keyboard, printer, USB, and LAN.
2
Remove the four M3 screws from the rear edge of the top panel. When
replacing, use a Torx T10 to torque to 5 in-lbs.
3
Slide the top panel back 1 to 2 inches and then lift it off.
4
To remove the bottom panel, flip the oscilloscope upside down and remove the
three M3 screws from the rear of the bottom panel.
5
Remove the remaining 13 screws from the bottom panel and lift the bottom
panel off the chassis.
Figure 1
Figure 2