Replacing Assemblies
Removing and replacing the acquisition boards/backplane assembly
1
Disconnect the power cable and remove the cover, top plate, and bottom plate.
2
Remove the motherboard and bulk +12 V power supply.
3
Turn the oscilloscope on its side and remove the SMAs connected to the
bottom of the backplane board.
4
Disconnect the BNCs from the rear of the instrument using a 9/16-inch open-
ended or socket wrench.