Troubleshooting
4
90000 A-Series Service Guide
71
Setting Up the BIOS
If the BIOS settings become corrupt, the Infiniium oscilloscope PC motherboard
will not recognize the hard drive and the oscilloscope may not start. To configure
the motherboard BIOS parameters to the default settings, follow these steps:
1
Connect the power cable to the oscilloscope.
2
Connect the external keyboard to the rear panel.
3
Press
[Delete]
as soon as the following prompt appears on the bottom of the
screen:
Press DEL to enter SETUP, F12 to select boot device.
If the prompt does not appear, or the oscilloscope does not appear to be
functioning, check the J1201 and J1202 motherboard power cable connectors.
Otherwise, continue with the next step.
4
Go to F3 Optimized Defaults and press
[Enter]
. Select Yes to load the defaults,
then press
[Enter]
.
5
Go to the Save and Exit Setup option and press
[Enter]
,
to save and exit the
setup. Select Yes, then press
[Enter]
.
Summary of Contents for Infiniium 90000A Series
Page 1: ...Keysight Infiniium 90000 A Series Oscilloscopes Service Guide...
Page 6: ...6 90000 A Series Oscilloscopes Service Guide...
Page 12: ...12 90000 A Series Service Guide 1 General Information...
Page 18: ...18 90000 A Series Service Guide 2 Calibration...
Page 116: ...116 90000 A Series Service Guide 5 Replacing Assemblies...
Page 120: ...120 90000 A Series Service Guide 6 Replaceable Parts Fan and Acquisition Assembly...
Page 121: ...Replaceable Parts 6 90000 A Series Service Guide 121 Power Supply and PC Motherboard...
Page 122: ...122 90000 A Series Service Guide 6 Replaceable Parts Sleeve and Accessory Pouch...
Page 128: ...128 90000 A Series Service Guide 7 Theory of Operation Oscilloscope block diagram...
Page 135: ......