Specifications and Characteristics
The following table contains a partial list of specifications and characteristics for
the Keysight Infiniium 90000 A-Series oscilloscopes. For a complete list, see the
data sheet at
www.keysight.com/find/90000
.
Specifications that are pertinent to each test are in Chapter 3, “Testing
Performance”. Specifications are valid after a 30-minute warm-up period, and
within
±
5
°
C from the temperature at which the last self-calibration was
performed.
Environment
Indoor use only
Ambient temperature
Operating: 5
°
C to +40
°
C
Non-operating: –40
°
C to +65
°
C
Humidity
Operating: up to 95% relative humidity (non-condensing) at +40
°
C
Non-operating: up to 90% relative humidity at +65 °C
Altitude
Maximum operating: up to 4,000 meters (12,000 feet)
Installation Category
II
Power Requirements
100-240 V, 47-63 Hz; 800 W max for 4-channel models
100-240 V, 50 Hz; 100-208 V, 50/60 Hz; 800 W max for 8-channel
models
Voltage Fluctuations
The mains supply voltage fluctuations are not to exceed ± 10% of the
nominal supply voltage
Pollution Degree
The Infiniium 90000 A-Series oscilloscopes may be operated in
environments of Pollution Degree 2
Pollution Degree Definitions
Pollution Degree 1: No pollution or only dry, non-conductive
pollution occurs. The pollution has no influence. Example: A clean
room or climate-controlled office environment.
Pollution Degree 2. Normally only dry non-conductive pollution
occurs. Occasionally a temporary conductivity caused by
condensation may occur. Example: General indoor environment.
Pollution Degree 3: Conductive pollution occurs, or dry,
non-conductive pollution occurs which becomes conductive due to
condensation which is expected. Example: Sheltered outdoor
environment.
Summary of Contents for Infiniium 90000A Series
Page 1: ...Keysight Infiniium 90000 A Series Oscilloscopes Service Guide...
Page 6: ...6 90000 A Series Oscilloscopes Service Guide...
Page 12: ...12 90000 A Series Service Guide 1 General Information...
Page 18: ...18 90000 A Series Service Guide 2 Calibration...
Page 116: ...116 90000 A Series Service Guide 5 Replacing Assemblies...
Page 120: ...120 90000 A Series Service Guide 6 Replaceable Parts Fan and Acquisition Assembly...
Page 121: ...Replaceable Parts 6 90000 A Series Service Guide 121 Power Supply and PC Motherboard...
Page 122: ...122 90000 A Series Service Guide 6 Replaceable Parts Sleeve and Accessory Pouch...
Page 128: ...128 90000 A Series Service Guide 7 Theory of Operation Oscilloscope block diagram...
Page 135: ......