Keysight E5260/E5270 Programming Guide, Edition 4
4- 123
Command Reference
Command Reference
*TST?
The *TST? query command performs the self-test and self-calibration, and returns
the results in ASCII format. Modules that fail the self-test are disabled, and can only
be enabled by the RCV command.
After the *TST? command, read the results soon.
Execution
Conditions
No module may be in the HIGH VOLTAGE state (forcing more than ±42 V, or
voltage compliance set to more than ±42 V).
To perform the self-test correctly, the measurement terminals should be opened.
Syntax
*TST? [
slotnum
]
Parameters
slotnum
:
Specifies the module to test. Integer expression. 0 to 9.
0: Mainframe and all modules. Default setting.
1 to 8: Module installed in the slot specified by
slotnum
.
9: Mainframe.
For HPSMU that uses two slots, set the greater slot number. For example, if it is
installed in slot 3 and 4,
slotnum
must be 4.
If
slotnum
specifies the slot that installs no module, this command causes an error.
Query Response
results
<CR/LF^EOI>
results
returns the sum of the following values corresponding to the failures.
Example
Statements
OUTPUT @E5270;"*TST?"
ENTER @E5270;A
results
Description
results
Description
0
Passed. No failure detected.
16
Slot 5 module failed.
1
Slot 1 module failed.
32
Slot 6 module failed.
2
Slot 2 module failed.
64
Slot 7 module failed.
4
Slot 3 module failed.
128
Slot 8 module failed.
8
Slot 4 module failed.
256
Mainframe failed.
Summary of Contents for E5260 Series
Page 1: ...Keysight Technologies E5260 E5270 Series of Parametric Measurement Solutions Programming Guide...
Page 3: ...User s Guide B2900 9001x xxxx 201x Edition x...
Page 13: ...1 Programming Basics...
Page 63: ...2 Remote Mode Functions...
Page 108: ...2 46 Keysight E5260 E5270 Programming Guide Edition 4 Remote Mode Functions Initial Settings...
Page 109: ...3 Programming Examples...
Page 185: ...4 Command Reference...
Page 335: ...5 Error Messages...
Page 349: ......